Goldman Halothane Vaporizer offered is made available by us in
without stand finish. For delivering efficient and precise functional
support in defined processes, the vaporizer system comes with provision of
vaporizing chamber available in form of glass bowl or bottle that contains
liquid anesthetic as well as a tap which helps in controlling proportion
of gas that flows through it. Being drawover type vaporizer, its precision
designs limit maximum concentration of Halothane vapor to 3% by volume.
This is achieved irrespective of total gas flow provided through unit.
Further, the system also allows adjustment through means of ratchet type
control valve. The unit also features glass jar being screwed on body and
allows quick detachment for easy filling. Further, the vaporizer is also
suitable with Trichloroethylene in low concentration.
Change Over Units offered is made available by us in
precision designed and constructed finish and finds usage in Anesthesia
Gas Apparatus. These change over units come normally positioned at top of
apparatus table and feature built-in emergency oxygen push button as well
as provision of change over facility from open to closed circuit as well
as from closed to open position. Here, we ensure the offered products are
made available in precision design and construction standards so as to
best match up with the defined operational demands.
Preset Regulators offered comprises Rao's Preset
Regulator that is made available by us in different operational finishes
including in meeting the handling demands of oxygen & Nitrous Oxide
gases used in anesthesia applications. Designed and developed for
anesthesia apparatus only, these come with outlets featuring non-return
valves so as to ensure there is not back flow of gases. Further, safety
valve is also incorporated in unit, thus avoiding need of safety valve in
tubing between regulator and rotameter unit. The models offered include
3625 - Oxygen and 3626 - Nitrous Oxide.